CLABUser’sGuide|23
sensorsbeleftto soakinacidsofbetween1‐3Mconcentrationforlonger
than48hours.TheexceptionstothisareHydrochlo ricacidHCLandSulphuric
acidH
2
SO
4
.Themaximumexposuretimefortheseacidsare:
Acid Maximum
Exposure Time
1 M HCL
2 M HCL
3 M HCL
20 minutes
10 minutes
5 minutes
1 M H
2
SO
4
2 M H
2
SO
4
3 M H
2
SO
4
48 hours
20 minutes
10 minutes
2.Suggestedexperiments
TheTemperaturesensorcanbeusedinmanyexperimentslike:
Monitoringindoorandoutdoortemperatures.
Monitoringfreezingandboilingwater.
Monitoringendothermicandexothermicreactions.
Specificheatexperiments.
Insulationinvestigations.
Solarenergystudies.
3.Technicaldata
Temperaturerange
40C..140C
Measuringelement
20kNTCThermistor; nonlinear
Calibration
(SteinhartHartequation)
Coefficients
T=[K
0
+K
1
*ln(R)+K
2
*(ln(R))
3
]
1
WhereTthetemperatureinC,and
Rthemeasuredresistancein
K
0
=1.02119E03;K
1
=2.22468E04;K
2
=1.33342E07
Maximumtemperaturethatsensor
cantoleratewithoutdamage
150C
Accuracy
2Cat‐40C
0.6at30C
1.8Cat140C
Responsetime
(timefor90%changeinreading)
10seconds(inwaterwithstirring)
270seconds(instillair)
80seconds(inmovingair)
Vue de la page 22